Audiogram form (1 pad of 50, 10 pads per case)
AM 232 Manual Audiometer Technical Specifications
Standards
The AM 232 meets ANSI S3.6 and IEC 645 Type 4 Audiometer Standards, and IEC 601-1, CSA C22.2 No. 601-1, and UL 544 electrical standards.
Frequency Range
Discrete Frequencies: 125, 250, 500, 750, 1000, 1500, 2000, 3000, 4000, 6000, 8000 Hz
Accuracy: ±3%
Total Harmonic Distortion: <2%
Intensity
Ranges:
125 Hz, -10 to 50 dB HL
500 to 6000 Hz, -10 to 90 dB HL
250 and 8000 Hz, -10 to 70 dB HL
(increments of 5 dB steps)
NOTE A “+10” dB switch extends maximum at all frequencies by 10 dB.
Accuracy: 125 to 4000 Hz, ±3 dB
6000 to 8000 Hz, ±5 dB
Signal to Noise Ratio: >70 dB
Tone Type
Rise/Fall Time: 20 to 50 msec
Continuous: Steady when present bar depressed
Pulse: 2.5 pulse/Sec
FM: 5 Hz, ±5%
Headset
Telephonics TDH39 Earphones with MX41AR Cushions (60 ohm impedance)
Power
Line Voltage: 100 to 120 V or 200 to 250 V (±10%)
Range: 50 to 60 Hz (±5%)
Consumption: 9 Watts
Battery
Types: Rechargeable (NiCad) or Non-Rechargeable (Alkaline)
Range: 7.0 V to 9.0 V
Capacity: NiCad—22 hours continuous operations, Alkaline—45 hours continuous operations
Mechanical
Dimensions: 13.25" W x 14" D x 3.75" H (33.66 cm x 35.56 cm x 9.53 cm)
Weight: 5.6 lb (2.53 kg) net 8 lb (3.64 kg) shipping 10 lb (4.55 kg) shipping with battery included
Additional Documentation
To view PDF documents, you need the free Adobe Acrobat Reader. If you do not already have it, you can download and install it directly from the Adobe website. When you are done with the PDF file, close the PDF window to return to this page.